Comes in the form of a cardboard sheet coated with a reactive purple ink that turns green when correct sterilisation conditions of time, temperature and saturated dry steal are met at particular sterilisation cycles.
![]() |
![]() |
The colour change of Steripoint® is inhibited by a mix of steam /air or gas, or when one of the parameters of the cycle is deficient.
The reactive ink turns from purple to green when correct sterilisation conditions of :
![]() |
![]() |
![]() |
![]() |
134° C 3.5 min | 134° C 5.3 min |
134° C 7 min |
134° C 18 min |
Box of 250 pieces Ref: 28635 Box of 2000 pieces Ref: 28035 |
Box of 250 pieces Ref: 28653 Box of 2000 pieces Ref: 28053 |
Box of 250 pieces Ref: 28607 Box of 2000 pieces Ref: 28007 |
Box of 250 pieces Ref: 28618 Box of 2000 pieces Ref: 28018 |
The reactive ink of Steripoint® is designed to change color when exposed to certain critical variables:
time, temperature and the presence of saturated steam for a specified sterilization cycle e.g : 134°C-7min / 121°C-20min, etc.
A constant reproducibility.
A quality commitment due to numerous quality checks of reactive ink colour changing during the manufacturing process.